Author(s):
Stefania Blasi1, Maurizio Levantino1, Luca Paperini2, Maria Grazia Bongiorni2 and Uberto Bortolotti1*
1Section of Cardiac Surgery, Italy
2Cardiology, Cardiac Thoracic and Vascular Department, University Hospital, Pisa, Italy
Received: 21 December, 2016; Accepted: 08 February, 2017; Published: 09 February, 2017
U Bortolotti, Section of Cardiac Surgery, University of Pisa, Via Paradisa, 2, 56124 Pisa, Italy, Tel: 39-50-995250; Fax: 39-50-541647; E-mail:
Blasi S, Levantino M, Paperini L, Bongiorni MG, Bortolotti U (2017) Extraction of an Infected Active Fixation Coronary Sinus Lead with the Aid of a Tissue Stabilizer. J Cardiovasc Med Cardiol 4(1): DOI: 10.17352/2455-2976.000036
© 2017 Blasi S, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Cardiomyopaties dilated; Endocarditis; Cardiopulmonary bypass
We present a 49-year-old man who required extraction of a cardiac resynchronization device due to lead infection. While 2 leads were successfully retrieved by a transvenous approach the third one with active fixation and entrapped into a coronary sinus side-branch, required surgical intervention. During the operation, performed on cardiopulmonary bypass on a beating heart, lead extraction was greatly facilitated by employment of a tissue stabilizer routinely used for off-pump coronary surgery.
Introduction
Coronary sinus (CS) lead implantation for cardiac resynchronization therapy (CRT) is associated with a significant rate of lead dislodgment [1,2]. The use of a lead with an active-fixation mechanism (e.g. deployable polyurethane lobes on the lead body) inserted into a CS branch allows reduction of lead dislodgment rate and decreases the risk of loss of pacing and phrenic nerve stimulation [3]. Because of fibrotic adhesions between polyurethane lobes and coronary sinus endothelium removal of implanted active-fixation leads, at variance with those with passive-fixation, may be technically challenging and sometimes it cannot be performed transvenously thus requiring surgical intervention [4].
We describe the case of a patient with an infected CRT device who required surgical removal of an active-fixation lead, entrapped into a CS collateral vein on cardiopulmonary bypass and beating heart; lead extraction was greatly facilitated by the use of a tissue stabilizer.
Case Report
A 49-year-old male was admitted to our hospital in October 2015 to perform a lead extraction procedure for a CRT device-related infection. In 2008, after evidence of dilated cardiomyopathy he received a CRT defibrillator employing three different leads among which an Attain StarFix™ Model 4195 (Medtronic Inc., Minneapolis, MN, USA) was implanted; this is an unipolar lead with an extendable active fixation mechanism through small polyurethane lobes which, after deployment, are conceived to firmly fixate the lead into a CS venous collateral. The patient had a good clinical response to the CRT therapy but after 5 years he had a reintervention to fix a traumatic fracture of the StarFix™ lead; after 14 months signs of local pocket infection appeared without signs or symptoms of systemic infection (white cells 7660/uL, ESR 12 mm/hr, procalcytonin 0,10 ng/ml). Hospital admission was performed to extract the entire CRT system. The generator was removed and transvenous extraction of the leads was attempted. Two of the three leads were completely removed with conventional mechanical dilatation [4], however, multiple attempts to remove the StarFix™ lead were unsuccessful since it appeared to be entrapped into a CS side-branch most likely occluded by thrombosis (Figure 1). Therefore surgical extraction of the lead was considered indicated.
- Gras D, Bocker D, Lunati M, Wellens HJ, Calvert M, et al. (2007) Implantation of cardiac resynchronization therapy systems in the CARE-HF trial: Procedural success rate and safety. Europace 9: 516-522. Link: https://goo.gl/GJ52hS
- McAlister FA, Ezekowitz J, Hooton N, Vandermeer B, Spooner C, et al. (2007) Cardiac resynchronization therapy for patients with left ventricular systolic dysfunction. A systematic review. JAMA 297: 2502-2514. Link: https://goo.gl/evhPqO
- Maytin M, Carrillo RG, Baltodano P, Schaerf RH, Bongiorni MG, et al. (2012) Multicenter experience with transvenous lead extraction of active fixation coronary sinus leads. PACE 35: 641-647. Link: https://goo.gl/FkTGk1
- Bongiorni MG, Di Cori A, Zucchelli G, Segreti L, De Lucia R, et al. (2011) A modified transvenous single mechanical dilatation technique to remove a chronically implanted active-fixation coronary sinus pacing lead. PACE 34: e66-e69. Link: https://goo.gl/nqErxQ
- Bongiorni MG, Zucchelli G, Soldati F, Arena G, Giannola G, et al. (2007) Usefulness of mechanical transvenous dilatation and location of areas of adherence in patients undergoing coronary sinus lead extraction. Europace 9: 69-73. Link: https://goo.gl/Cti3S1
- Cronin EM, Ingelmo CP, Rickard J, Wazni OM, Martin DO, et al. (2013) Active fixation mechanism complicates coronary sinus lead extraction and limits subsequent reimplantation targets. J Interv Card Electrophysiol 36: 81-86 Link: https://goo.gl/fOyI8a
Figure 1:
The procedure was performed through a standard median sternotomy; an intraoperative trans-esophageal echocardiogram showed a 30% ejection fraction with moderate mitral regurgitation. Due to the hemodynamic instability during cardiac manipulation it was decided to perform lead extraction on pump with a beating heart. The ascending aorta and the right atrium were cannulated and normothermic cardiopulmonary bypass instituted. The heart was then lifted upwards and on the right side exposing the distal part of the lead which was entrapped into the posterior left ventricular coronary vein. To facilitate lead extraction an Octopus Tissue Stabilizer® (Medtronic Inc., Minneapolis, MN, USA) was used for focal immobilization of the ventricular wall; a 4 cm incision of the vein confirmed vein thrombosis and exposed the lead which was isolated and could be removed only after dissection of the fixing polyurethane lobes which were incorporated in the vein wall (Figure 2). A new bipolar drug-eluting MyoDex 1084T epicardial lead (St. Jude Medical, Saint Paul, MN, USA) was screwed in the lateral wall of the left ventricle and tunnelled in the right prepectoral region for the connection to a new CRT device. The patient was weaned from bypass without difficulty and the post-operative course was uneventful.
Figure 2:
Discussion
The Attain Starfix™ Model 4195 is a steroid eluting unipolar CS lead with an active-fixation mechanism by means of three soft polyurethane lobes near its tip which allow the lead to be fixed once positioned into a CS collateral vein. While deployment and removal of such lead has been reported to be achieved successfully in the acute setting [3], information regarding both function of the fixation mechanism and easiness of lead removal in the long-term are limited [3]. As shown previously [5] transvenous extraction of the StarFix™ lead may be extremely challenging requiring complex mechanical dilatation techniques in order to completely remove it without fracturing and dislodgement of the polyurethane lobes. Furthermore, as showed by venography also in the present case, this particular lead is often associated with a significant thrombotic reaction with chronic occlusion of the CS vein branch and lead entrapment [5]. Cronin et al. described CS vein occlusion and fibrotic adhesions between polyurethane lobes and CS endothelium as an important mechanism making transvenous lead extraction extremely difficult and preventing implantation of a new lead in the same CS vein branch [6].
Many techniques to perform transvenous lead removal have been proposed, ranging from isolated direct manual traction to use of locking stylets, to cutting or laser extraction sheaths [6]. However, when transvenous lead extraction results unsuccessful surgical intervention is necessarily required. In this respect, Maytin et al. have reported a thoracoscopic approach, through a left anterior and mid axillary access, for surgical extraction of a CS lead [3]. In the present case extraction of a CRT device was mandatory because of infection. Removal of the endocardial leads by a transvenous approach resulted in failure to extract one active-fixation lead entrapped into a CS collateral. Surgical extraction was therefore required which was performed with the aid of cardiopulmonary bypass without arresting the heart, due to the need of heart manipulation in the presence of a dilated and hemodynamically unstable left ventricle. Operation was greatly facilitated by employment of a tissue stabilizer, commonly used during off-pump myocardial revascularization procedures; this provided a quiet operative field, allowing complete removal of the lead together with its polyurethane lobes thus contributing to eradicate the infection and furthermore avoided the need for cardioplegic arrest.
The present case describes a potential complication of an active-fixation pacing lead suggesting a further use of the Octopus TissueStabilizer®.